User Contributed MET/CAL PROCEDURE ============================================================================= INSTRUMENT: Kyowa Wow & Flutter Tape Speed Sub DATE: 27-Aug-97 AUTHOR: User Contributed REVISION: ADJUSTMENT THRESHOLD: 70% NUMBER OF TESTS: 16 NUMBER OF LINES: 131 CONFIGURATION: HP 34401A STANDARD: Panasonic VP-7750P ============================================================================= STEP FSC RANGE NOMINAL TOLERANCE MOD1 MOD2 3 4 CON 1.001 STD Panasonic VP-7750P 1.002 DISP Set the Wow & Flutter Meter as follows: 1.002 DISP OSC OUT ............................ TEST 1.002 DISP INDICATION ......................... CCIR 1.002 DISP METER RANGE % ...................... 3 1.002 DISP FUNCTION ........................... LIN 1.003 DISP Connect the 34401A INPUTS to the Wow & Flutter Meter 1.003 DISP TAPE SPEED output. 1.004 DISP Connect the Wow & Flutter Meter OSC OUTPUT 1.004 DISP to the UUT CH 1 RECORD INPUT BNC. 1.005 DISP Connect the UUT OUTPUT 1 BNC to the Wow & Flutter 1.005 DISP Meter INPUT. 1.006 DISP On the Wow & Flutter Meter adjust the CCIR/DIN 1.006 DISP TEST ADJ control. 1.007 DISP Set the Wow & Flutter Meter as follows: 1.007 DISP OSC OUT ............................ 3kHz 1.008 HEAD {} 1.009 HEAD { 38 cm/s } 1.010 JMP 2.001 1.011 EVAL 2.001 IEEE [@34401]MEAS:VOLT:DC?[I] 2.002 MEME >10 2.003 DISP Using the UUT TAPE SPEED UP button, set the UUT to 2.003 DISP 38 cm/s. 2.004 DISP On the UUT depress the FWD and REC buttons 2.004 DISP simultaneously. 2.005 DISP On the Wow & Flutter Meter change the METER INDICATION % 2.005 DISP range until an onscale reading is available. 2.006 MEMI Enter the Wow & Flutter meter indication in %. 2.007 ACC 0.000%p-p 0.015U 2.008 MEME 2.009 MEMC %p-p -0U +0.300U w&f 3.001 IEEE [@34401][D500]MEAS:VOLT:DC?[I] 3.002 MATH MEM = (MEM - M[10]) / .77 3.003 ACC .000pct 1.4e-6U 3.004 MEME 3.005 MEMC pct 0.2U speed 4.001 HEAD {} 4.002 HEAD { 19 cm/s } 4.003 JMP 5.001 4.004 EVAL 5.001 DISP Press STOP on the UUT. 5.002 DISP Using the UUT TAPE SPEED UP button, set the UUT to 5.002 DISP 19 cm/s. 5.003 DISP On the UUT depress the FWD and REC buttons 5.003 DISP simultaneously. 5.004 DISP On the Wow & Flutter Meter change the METER INDICATION % 5.004 DISP range until an onscale reading is available. 5.005 MEMI Enter the Wow & Flutter meter indication in %. 5.006 ACC 0.000%p-p 0.015U 5.007 MEME 5.008 MEMC %p-p -0U +0.3U w&f 6.001 IEEE [@34401]MEAS:VOLT:DC?[I] 6.002 MATH MEM = (MEM - M[10]) / .77 6.003 ACC .000pct 1.4e-6U 6.004 MEME 6.005 MEMC pct 0.2U speed 7.001 HEAD {} 7.002 HEAD { 9.5 cm/s } 7.003 JMP 8.001 7.004 EVAL 8.001 DISP Press STOP on the UUT. 8.002 DISP Using the UUT TAPE SPEED UP button, set the UUT to 8.002 DISP 9.5 cm/s. 8.003 DISP On the UUT depress the FWD and REC buttons 8.003 DISP simultaneously. 8.004 DISP On the Wow & Flutter Meter change the METER INDICATION % 8.004 DISP range until an onscale reading is available. 8.005 MEMI Enter the Wow & Flutter meter indication in %. 8.006 ACC 0.000%p-p 0.015U 8.007 MEME 8.008 MEMC %p-p -0U +0.4U w&f 9.001 IEEE [@34401]MEAS:VOLT:DC?[I] 9.002 MATH MEM = (MEM - M[10]) / .77 9.003 ACC .000pct 1.4e-6U 9.004 MEME 9.005 MEMC pct 0.2U speed 10.001 HEAD {} 10.002 HEAD { 4.8 cm/s } 10.003 JMP 11.001 10.004 EVAL 11.001 DISP Press STOP on the UUT. 11.002 DISP Using the UUT TAPE SPEED UP button, set the UUT to 11.002 DISP 4.8 cm/s. 11.003 DISP On the UUT depress the FWD and REC buttons 11.003 DISP simultaneously. 11.004 DISP On the Wow & Flutter Meter change the METER INDICATION % 11.004 DISP range until an onscale reading is available. 11.005 MEMI Enter the Wow & Flutter meter indication in %. 11.006 ACC 0.000%p-p 0.015U 11.007 MEME 11.008 MEMC %p-p -0U +0.5U w&f 12.001 IEEE [@34401]MEAS:VOLT:DC?[I] 12.002 MATH MEM = (MEM - M[10]) / .77 12.003 ACC .000pct 1.4e-6U 12.004 MEME 12.005 MEMC pct 0.2U speed 13.001 HEAD {} 13.002 HEAD { 2.4 cm/s } 13.003 JMP 14.001 13.004 EVAL 14.001 DISP Press STOP on the UUT. 14.002 DISP Using the UUT TAPE SPEED UP button, set the UUT to 14.002 DISP 2.4 cm/s. 14.003 DISP On the UUT depress the FWD and REC buttons 14.003 DISP simultaneously. 14.004 DISP On the Wow & Flutter Meter change the METER INDICATION % 14.004 DISP range until an onscale reading is available. 14.005 MEMI Enter the Wow & Flutter meter indication in %. 14.006 ACC 0.000%p-p 0.015U 14.007 MEME 14.008 MEMC %p-p -0U +0.6U w&f 15.001 IEEE [@34401]MEAS:VOLT:DC?[I] 15.002 MATH MEM = (MEM - M[10]) / .77 15.003 ACC .000pct 1.4e-6U 15.004 MEME 15.005 MEMC pct 0.2U speed 16.001 END